Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine

Product Details
Customization: Available
Certification: CE
Test Item: Protein, Glucose, Blood, pH
Diamond Member Since 2022

Suppliers with verified business licenses

Audited Supplier Audited Supplier

Audited by an independent third-party inspection agency

Year of Establishment
2022-04-28
Plant Area
220 square meters
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
  • Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine
Find Similar Products
  • Overview
  • Product Description
  • Detailed Photos
  • Company Profile
Overview

Basic Info.

Model NO.
SY-V2809
Product Name
Visual Electro Electrophysiology System
Usage
Human/ Animal
Input Impedience
≥ 1000 MΩ
Cmrr
≤ 4μvpp
a/D Resolution
24 Bit
Luminance
0-200 CD/M2
Contrast
> 98%
Stimuli Frequency
0.01-200Hz
Monitor
Infra-Red Camera
Transport Package
Carton Box
Trademark
Sunnymed
Origin
China
Production Capacity
200 Set/Month

Product Description

SY-V2809 Recording System Ophthalmological visual Electrophysiology Measurement Machine

Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine

 
Product Description
 
Amplifier
 
Input Impedance
≥ 1000 MΩ
CMRR
≥ 115dB
Input Noise
≤ 4uVpp
Filter
Full real-time digital filter, band-pass, band-stop, high-pass,
low-pass, comb filter applies to every channel independently
A/D Resolution
24 bit
Port
Optical isolated USB 2.0
Power Supply
Li-ion battery, switch to USB power automatically
when low battery
Flash Stimulator
 
Light Source Color
White, Red, Green, Blue, Yellow and any of their combination
Flash Luminance
Can produce 30cd·s/m2 broad-band white flash in < 4ms
Flash Luminance Range
0-100 cd·s/m2
Stimuli Mode
Single, Double, String, On/Off, On duration can be adjusted
Background Illumination
0-500 cd/m2
Stimuli Frequency
0.01-200Hz
Fixation & EOG
3 Fixation/EOG indicator, 64 adjustable intensity;
Monitor
Infra-red camera
Pattern Stimulator
 
Luminance
0-200 cd/m2
Contrast
> 98%
Pattern Shape
Checkerboard, Horizontal/vertical stripes, rectangle,
adjustable element angle
Stimuli Field
Full field, left/right/upper/lower half, up-down/left-right margin;
¼ left-top/right-top/left-bottom/right-bottom;
horizontal/vertical center, 1/8 center;
Center circle with any angle;
Field outside center circle with any angle.
 
 
Detailed Photos
 

 

Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement MachineSy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine

 
Company Profile

 

 
Sy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement MachineSy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement MachineSy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement MachineSy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement MachineSy-V2809 Recording System Ophthalmological Visual Electrophysiology Measurement Machine

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now