Customization: | Available |
---|---|
Application: | Hospital, Factory, Office |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
SpO2 Module (Options):
Measurement Range: 30~100% ±2% between 80%~90% others ±5%
Alarm Range: User set high limit and low limit.
Alarm Accuracy: ±10% within setting values.
Alarm time Accuracy: less than 12 sec.
Pulse Oximeter
NIBP (Options):
Method: automatic oscillation
Working mode: manual/automatic
Measure: adult: 10-250mmHg
Ement: pediatric: 10-200mmHg
Rage: neonatal: 10-135mmHg
EtCO2 Module (Options):
CO2 Range 0 to 13%(0-98.8mmHg@BTPS)
CO2 Accuracy <5.0% CO2 (at BTPS) ± 3 mmHg
>5.0% CO2 (at BTPS) < 10% of reading
Breath Rate 3~60 BPM
Measurement Accuracy ±1BPM
Pacemaker (Options):
Type: VVI
Pulse Type: rectangular,constant current
Pulse Amplitude: 0 to 180 mA ± 10% or 5 mA (whichever is greater).
Pacing Rate: Variable from 30 ppm to 180 ppm ±1.5% (increments or decrements by a value of 2 ppm)
Multi-Function Electrode: (MFE) Pads: multipurpose defibrillation/pacing electrodes
Pause: Pacing pulse frequency reduced by a factor of 4 when activated
Refractory Period: NC(VVI demand do not have this specification)
Pulse Width: (20+ 1.5 ms)
Output Protection: Fully defibrillator protected and isolated
Recorder
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.
AED Mode
AED Function: Auto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Manual Mode
Energy Selection: Selectable at 2 5 7 10 20 30 50 70 100 150 200 300 360 joules.
Synchronized Mode: Synchronizes defibrillator pulse to patient's
R-wave. "SYNC" message displayed on monitor.
Standard Configuration:
ECG Cable (5 Leads) ...... 1Pcs
AED Pad Adapter ........... 1Pcs
AED Pad ..........................1pcs
Ground Cable ................ 1Pcs
Power Cable ................... 1Pcs
Thermal Paper Role ........ 1Pcs
Cleaning Cloth ................ 1Pcs
Patient connection
|
3-lead ECG cable or 5-lead ECG cable paddles
|
||
Lead selection
|
Display on monitor paddles | ll | AVR AVL AVF V.
|
||
ECG size
|
0.25 0.5 1 1.5 2 4 cm/mV displayed on the monitor.
|
||
AED function
|
automatic analysis and charging x3 with programmable automatic energy level selection. on-screen prompts and voice prompts
|
||
Type of charging prompts
|
Voice and visual prompts
|
||
Charge control
|
front panel, push button on paddle.
|
1. How to buy?
2. How long to wait for the order?
A. Ready stock within 24 hours
B. OEM need 2-7 working days
C. Safety delivery:special one-to-one inspection service before arranging the delivery
3. How to use our product?
A. Video tutorial: including installation, operation, repairing and maintenance
B. Real Time Online Video Call at your convenience: Pre-sales and after-sales online video training
4. Warranty and after-sales service
A.12-60 months
B. After-sales service: Reply within 8 hours