Customization: | Available |
---|---|
Application: | Hospital, Factory, Office, Public Place |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Output
|
Biphasic Truncated Exponential
|
Energy Sequence
|
150, 150, 200J
|
Charge Time
|
8 sec. to 150J; 10 sec. to 200J (New Battery)
|
Analysis Time
|
9 sec.
|
The maximum time from the initiation of rhythm analysis to
readiness for discharge with a new battery |
Less than 30 seconds |
The maximum time from the initiation of rhythm analysis to
readiness for discharge after 6 shocks |
Less than 35 seconds
|
The maximum time from initially switching power on to
readiness for discharge |
Less than 25 seconds
|
Audible Prompts
|
22 audible prompts
|
Visual Prompts
|
LED prompts
|
Controls
|
Two buttons-ON/Off, Shock
|
Output Energy Accuracy
|
±15% into any impedance from 25 to 175Ω
|
The Maximum Voltage
|
1200±50V
|
Output disabled when the patient impedance is outside limits
|